logo

Detalhes dos produtos

Created with Pixso. Para casa Created with Pixso. produtos Created with Pixso.
Controlador de ESP
Created with Pixso.

Fluke 1524 Reference Thermometer | Industrial Temperature Calibration

Fluke 1524 Reference Thermometer | Industrial Temperature Calibration

Nome da marca: Fluke
Número do modelo: 1524·
MOQ: 1
preço: Negotistical
Informações pormenorizadas
Lugar de origem:
OEM
Canais de entrada:
2
Detalhes da embalagem:
Nova caixa original
Descrição do produto
Fluke 1524 Reference Thermometer | Industrial Temperature Calibration
The Fluke 1524 Reference Thermometer delivers precision industrial temperature calibration with advanced measurement capabilities and comprehensive data recording.
Key Specifications
Specification Details
Input Channels 2
Data Recording 15,000 measurements with date and time stamps; 25 readings and statistics
Sampling Interval (Standard) 1 second
Sampling Interval (Fast Mode) 0.3 seconds (See technical manual for details)
Sensor Types PRT, RTD, Thermistor, and Thermocouple
Thermocouple Types C, E, J, K, L, M, N, T, U, B, R, S
Operating Temperature -10 ºC to 60 ºC (Optimal accuracy range 13 ºC to 33 ºC)
Power Requirements 3 AA alkaline batteries
Dimensions 96 x 200 x 47 mm (3.75 x 7.9 x 1.86 in)
Weight 0.65 kg (1.4 lb)
Technical Performance Specifications
Ambient Conditions for Optimal Accuracy: 13 °C to 33 °C
Millivolt Range and Accuracy: -10 mV to 75 mV ± (0.005% + 5 μV)
Resistance Range and Accuracy: 0 Ω to 400 Ω ±(0.004% + 0.002 Ω)
200 Ω to 50 kΩ ±(0.01% + 0.5 Ω)
50 kΩ to 500 kΩ ± (0.03%)
Temperature Coefficient, Voltage: (-10 °C to 13 °C, +33 °C to 60 °C) ±(0.001%/ °C + 1 μV/ °C)
Temperature Coefficient, Resistance: 0.0008%/ °C + 0.0004 Ω (0 Ω to 400 Ω)
0.002%/ °C + 0.1 Ω (0 Ω to 50 kΩ)
0.06%/ °C + 0.1 Ω (50 kΩ to 500 kΩ)
Excitation Current, Resistance: 1 mA (0 Ω to 400 Ω)
10 μA (0 Ω to 50 kΩ)
2 μA (50 Ω to 500 kΩ)